Analytical Lab High Purity Water Purification Systems

Analytical Lab High Purity Water Purification Systems
Model:-
Brand:ZHUOYUE
Origin:Made In China
Category:Industrial Supplies / Environment Protection Facilities / Other Environment Protection Facilities
Label:Lab water , ro water , water purifier
Price: ¥1110 / pc
Min. Order:1 pc
Inquire Now

Product Description

high purity water

 

Model ZYCGF-I-60L ZYCGF-II-60L ZYCGF-III-60L ZYCGF-IV-60L
Output(25°C)* 60 liters/hour
Flow rate Up to 1.8 liters/minute(with pressure tank)
Feed water requirements Tap water, temperature:5-45°C,pressure:1.0-4.0Kgf/cm2
RO water quality
Conductivity At 25℃ 2-10μs/cm
Ion rejection rate  99% (new RO membrane)
Organic rejection rate >99%, when MW>200 Dalton
Particles and bacteria rejection rate >99%
High Purity water quality
Resistivity(25°C) 18.2MΩ.cm
Heavy metal ion <0.1ppb
Bacteria < 1cfu/ml
Particle(>0.2μm) <1/ml
TOC Level*** < 20 ppb < 10 ppb
Dimension and weight Length×Width×Height:530×400×590mm, 40Kg
Electrical requirements AC210-240V, 50/60Hz
Power 30W-50W
Terminal disinfection filter  N/A yes
UV lamp N/A yes
UF N/A yes
Standard configuration N/A N/A yes
Can be selected flow rate  40L/H, 60L/H, 80L/H,100L/H

 

Typical Features of High Purity Water System:

1. Water Quality on-line monitoring; accurate and intuitive LCD panel digital display;

2. Polyethylene (PE) and polyacetal (POM) made pressure tank avoids secondary pollution;

3. Automatic water production and storage;

4. Produce RO pure water and UP ultra-pure water.

5. System flushing extends life span of reverse osmosis membrane 

6. Pressure alarm display.

7. Automatic water quality alarm.

Working Conditions of High Purity Water System:

Feed water have certain requirements. Feed water quality and environmental changes affect systems capacity. Failure to meet the requirements of feed water may result in contamination and damage to membrane components. 

Source water : city tap water or groundwater

Minimum water supply pressure: 0.1-0.5MPa (pressure at minimum feed water flow rate.)

Water temperature: 5 ~ 35 ℃

PH range: 6 ~ 9

Solubility Total solids: TDS ≤ 200 mg / L (ppm)

Residual chlorine: ≤0.1mg / L (ppm)

Total iron: ≤0.1 mg / L (ppm)

Manganese: ≤0.5 mg / L (ppm)

Co2: <30 mg / L (ppm)

SDI: <5 mg / L (ppm)

Note: TDS >350mg / L ,the optional water softener should be taken into consideration.

 

high purity water

Technical Process of Analytical Lab High Purity Water Purification Systems: high purity water

The technical process is divided into source water pretreatment system,reverse osmosis system and ultra-pure system.

1. Source water pretreatment system:  PP fiber filter + AC filter + Softener 

2. Reverse osmosis system: It has high removal rate more than 99% for high valent ions, colloids, bacteria and organic matter (including endotoxin) with molecular weight greater than 300Dalton.

3. Ultra pure system: Ion exchange resin using H + exchange cation out the water in the filter ,using OH- exchange anions out the water  , the replaced H + and OH- combined to produce H2O, ultra-purified column filling the United States Dow UPW-450 Ultra-pure water for nuclear grade resin, the water resistance of up to 18.2MΩ.cm.

Installation pictures of High Purity Water Purification Systems

high purity water

 

high purity water

Analytical Lab High Purity Water Purification Systems 1

Member Information

Sichuan Zhuoyue Water Treatment Equipment Co.,Ltd.
Country/Region:Si chuan - China
Business Nature:Manufacturer
Phone:13808197592
Contact:Monica Huang (General Manager)
Last Online:25 Apr, 2019