Model: | GF303D |
---|---|
Brand: | GFUVE |
Origin: | Made In China |
Category: | Electronics & Electricity / Electronic Instrument / Electricity Meter & Instrument |
Label: | power source , phantom load , three phase source |
Price: |
US $20008
/ pc
|
Min. Order: | 1 pc |
Last Online:02 Jun, 2021 |
It is suitable for power department, measuring department, quality control department, research units, and institutions of higher learning, and electric energy meter, power distribution terminals, power supply management, load control, power quality, reactive power compensation device and the production of enterprise etc.
Main Functions and Features
To set up and take the load regulation of voltage, current, phase angle ,frequency and power factor etc;
To set 2 ~ 31 times of the amplitude and phase harmonics, and it can be added to the base wave in every harmonic output;
Voltage, current output a wide range, big power, high stability, waveform distortion degree is small;
Strong with load ability, but it can take capacitive, sensibility, resistance of composite type load or load and load regulation is higher than 0.01%;
To apply the 32 bit MPU + DSP + CPLD, powerful flexible;
To use hardware PID, fast response, the change of load will not cause output volatility;
Industrial frequency waves as high as 5000 points a week of the wave kneading, signal output without filter, precise output waveform, the harmonic output precision, harmonic distortion degree is small;
Range switching automatically
To use software calibration, simple operation, stable and reliable;
Display of vector diagram, adjustable to IEC387 OR
To have big screen, 7-inch TFT color LCD display, Chinese or English interface, operating simply;
To have perfect over-current, over-voltage, heat, a shorts-and-opens, overload protection, failure detection automatically;
With RS232 interface, and PC connection
To take place machine software, it can be output through the PC software program;
Technical Data
AC voltage output
Range: 57.7 V, 100 V, 220 V, 380 V; range switch automatically (L1, L2, L3 phase)
Adjust fineness: 0.01% RG
Accuracy: 0.05% RG
Stability: <0.02% RG / 120 s
Distortion degree: <0.5% (not capacitive load)
Output power: 25 VA
Full load regulation rate: 0.01% RG
Full load regulation time: less than 1 mS
Long-term stability: ± 60 PPM/year
AC current output
Range:
Adjust fineness: 0.01% RG
Accuracy: 0.05% RG
Stability: <0.02% RG / 120 s
Distortion degree:< 0.5% (not capacitive load)
Output power: 45 VA
Full load regulation rate: 0.01% RG
Full load regulation time: less than 1mS
Long-term stability: ± 60 PPM/year
Power output
Accuracy: 0.05% RG
Stability: 0.01% RG / 120 s
Phase angle
Adjusting range: 0 ~ 359.99°
Resolution: 0.001°
Accuracy: 0.05°
Frequency
Adjusting range: 45Hz ~ 65Hz
Resolution: 0.001 Hz
Accuracy: 0.005 Hz
Temperature drift: ± 0.5 PPM / ℃
Long-term stability: ± 4 PPM/year
Power factor
Adjusting range:-1 ~ 0 ~ + 1;
Resolution: 0.0001;
Accuracy: 0.0005.
Harmonic accuracy
Harmonic phase: 0 ~ 359.99°
Harmonic phase accuracy: < 0.01°
Harmonic set accuracy: 0.1% (relative to the base wave ratings)
Power supply:
One phase AC 180~265 V, frequency 50/60 Hz
Environmental conditions:
1. Working temperature:
2. Relative humidity: ≤ 85% ;
3. Storage condition:
Volume and weight:
Size: