Model: | HLCS-10000 |
---|---|
Brand: | TE |
Origin: | Made In China |
Category: | Electronics & Electricity / Electronic Instrument / Electricity Meter & Instrument |
Label: | calibrate equipment , load cell calibrator , sensor calibration |
Price: |
-
|
Min. Order: | 1 pc |
Last Online:12 Apr, 2016 |
Model:ELCS-500
Description
It is specially designed for the calibration of load cells. It can calibrate the load cell within ±0.5%.
Advantage
Advantage of Electromechanical Type than Hydraulic calibration system
Load speed can be started from 0.001mm/min up to 250mm/min, while Hydraulic is from 0.1mm/min to 50mm/min
No oil leakage and maintenance free.
Note: this electromechanical calibration system can be customized upon request.
Specifications
Max. Load: 500kN
Deviation of load value: 0.05%
Load repeatability: 0.05%
Sensitive limit: 0.02%
Accuracy: ≤±0.0005%F.S
Stabilization time at any point: ≥30min
Load/unload time: any time set
Load speed: 0.001-250mm/min
Stiffness: 600kN/mm
Model:HLCS-10000
Description
It is specially designed for the calibration of load cells. It can calibrate the load cell within ±1%.
Main Structure
1. Load Frame
The load frame is with four columns. The columns are fastened by nuts. Such design is with high stiffness and no gap loading design.
Both tensile and compression are at one space. Max. Compression load is 10,000kN (1000t) and max. tensile load is 5,000kN (500t). The measuring load cell of INTERFACE 1060 is with measuring range 9000kN. And max. calibration space is 1100mm.
Specifications
Sampling rate: 1000Hz (EDC220); 5000Hz(EDC580)
Communication Processor, AMD520, 133MHz
Load resolution ±180,000 steps, two ranges 2mV/V and 4mV/V
RS485 Interface for external keyboard/Display supports only one device
Drive Interface ±10V (16bit) analogue command output OR digital command output
Safety functions like 4DI
Two Option slots for second analogue channel for strain gauge extensometer OR, two encoder channels for extensometer with one or two encoder.
PC communication via USB or Ethernet
Internal socket for serial Sensors (COM 1)
Internal socket for Debug (COM 2)