Salt Spray Chamber
Uses
This machine has a standardized test method to check corrosion resistence of coated samples. Salt spray test is an accelerated corrosion test that produces a corrosive attack to the coated samples in order to predict its suitability in use as a protective finish. The appearance of corrosion products (oxides) is evaluated after a period of time. Test duration depends on the corrosion resistance of the coating; the more corrosion resistant the coating is, the longer the period in testing without showing signs of corrosion. Different coatings have different behaviour in salt spray test and consequently, test duration will differ from one type of coating to another.
Character
1. the spray nozzle the heart of teser adopt Bo Nute Principle which change saline water into atomization but no crystallization produced.
2. the controller is digital type microcomputer which delivers precision temparature
3. parameters show on the display screen such as temparature, humidity, pressure, spray capacity and time.
4.test specifications are accept worldwild.
|
Specification
Design standard |
GB/T 10125-1997,ASTM.B117-97 JIS H8502 ,IEC68-2-11,IEC68-2-52 1996 GB.10587-89,CNS.4158,CNS.4159,CASS GB/T 12967.3-91 CASS |
|
Internal size |
90×60×50cm |
|
Temperature |
lab: NSS.ACSS35℃±1℃/ CASS50℃±1℃ |
|
Saturated air cylinder |
NSS.ACSS35℃±1℃/ CASS50℃±1℃ |
|
Spray capacity |
1~10cc/h |
|
Structure material |
F.R.P |
|
Accessories |
test salt, barrel, tool |
|
Power |
1φ、220V/50HZ |
Design Standards:ISO 9227,GB/T 2423.17-1993 ,GB/T 2423.18-2000,GB/T 10125-1997,ASTM.B117-97
JIS H8502 ,IEC68-2-11,IEC68-2-52 1996 ,GB.10587-89,CNS.4158,CNS.4159,CASS,GB/T 12967.3-91 CASS
1.Inner testing size:900×600×500(L×W×H)mm
2.Outside size:115×630×1185(L×W×H)mm
3.Power:220V,1phase, 3KW,50/60HZ
4.Material:
(A) Test room uses P.V.C (polrvinyl chloride) plate,thickness8mm,heat-resistant can reach to80℃。
(B) Sealed cap of test room uses imported organic glass,Thickness is8mm,heat-resistant can reach to80℃。
(C) Reagent replenishing bottle uses hidden water level perspective window,easy to clean,hard to broke down.
(D) Saturation air barrel uses platsic and high pressure resistant barrel to make sure heat preservation is good.And match with communicating vessels and cleaning devices.
(E) Shelves of test room uses complanate dividing shelf,can adjust the angle freely,spray at all sides,can put many test specimen.
5.CNS.JIS.ASTM specification
(A) Salt water spray test:NSS , ACSS
(1)Test Room:35℃±1℃
(2)Saturation air barrel :47℃±1℃
(B) Corrosion resistance test(acidity test): CASS
(1)Test Room:50℃±1℃
(2)Saturation air barrel:63℃±1℃
Cubage of the brine barrel :15L
Brine Concentration : 5%
6.Air supply system:air pressure is1kg/cm2 ,separate into 2 stage to adjust
First stage is roughly adjustment,pressure is2kg/cm2,Second is accurate adjustment,pressure is1kg/cm2
7.Spraying methods:
(A) Adopt Bernoulli's Principle,absorb the salt water and spray,spraying extent is uniform,don't have blocking and crystallizing function to make sure the continuation testing spray.
(B) Nozzle:Use special glass nozzle material,can adjust the size of spraying amount and the angle.
(C) Spray amount:1 pcs spray tower is putted in the centre of test room to make sure spraying uniform.It can adjust as 1-2ml 80cm2/hr.
(D) Chamber has demisting function,you can see the sample after testing.
8.Heating methods:Adopt direct heating methods,heating speed is quick,can reduce standby time,when temperature is reaching,you can use ON/OFF to change.Temperature can be setted from normal to70C.
9.Control system
1.Test room temperature controller 0-99℃,liquid digit showing with PID calculation function,controlling error ±0.1℃ 1PCS
2.saturated air barrel temperature controller 0-99℃,liquid digit showing with calculation function,controlling error±0.1℃ 1PCS
3.Heating trough liquid swelling safe temperature controller 0-120℃ 1PCS
4.Saturated air barrel swelling temperature controller0-120℃ 1PCS
5.Time showing controller:0.01s-9999hr 1PCS
6.Total Timer 0-9999hr 1PCS
7.Temperature controlling error(Test Room):±0.1℃
8.Temperature controlling error(Saturation air barrel):±0.1℃
9.Timer showing:0.1s-99990hr.
10.Accumulate total testing time automatically.
11.Showing relative humidity automatically.
10.Water system:Adopt water replenishing system,when the water level is too low,it will replenish automatically.
11.Safety protection device:
(A) Low water level,chamber will cut off power automatcially.
(B) Over-temperature,chamber will cut off power automatcially.
(C) With safety alarm light device.
(D) Have earth leakage protection device.
12.Accessories(Machine spare parts)
(A)Ltype commodity shelf 4pcs
(B)measuring cylinder 1pcs
(C)Temperature pointer 2pcs
(D)Otype commodity shelf 6pcs
(E)collector80cm 1pcs
(F)glass nozzle 1pcs
(G)Temperature cup 1pcs
(H)demist system 1pcs
(I)NACL 500g 2pcs
(J)English Manual and video 1pcs
Payment Terms: | TT |
---|---|