Model: | ATS-1000 |
---|---|
Brand: | ACE |
Origin: | Made In China |
Category: | Electronics & Electricity / Electronic Instrument / Laboratory Instrument |
Label: | Thermal Shock , Preheat room , Temperature shock |
Price: |
-
|
Min. Order: | 1 pc |
Last Online:28 Jul, 2020 |
1. Standard
Model |
ATS-162 |
ATS-340 |
ATS-500 |
ATS-1000 |
Workroom dimensions (cm) |
45*45*45 |
60*60*60 |
80*80*80 |
100*100*100 |
Exterior dimensions (cm) |
156*87*154.5 |
171*102*184.5 |
191*122*226.5 |
211*142*266.5 |
Sample zone dimensions |
30*30*25 |
45*45*36 |
65*65*50 |
85*85*70 |
Temperature range |
A:-20℃~200℃ B:-40℃~200℃ C:-60℃~200℃ |
|||
Temperature fluctuation |
≤±2℃ for high temperature and low temperature |
|||
Temperature fluctuation |
≤±0.5℃ (under constant temperature) |
|||
Sample zone load |
20kg |
30kg |
50kg |
60kg |
2. Customized
Preheat room |
Upper limit preheat temperature |
200℃ |
||
Temperature ramp up rate |
R.T.~+200℃, about 40min (unload) |
|||
Precool room |
Lower limit precool temperature |
-75℃ |
||
Temperature ramp down rate |
R.T.~-70℃, about 90min(unload) |
|||
Workroom |
Temperature shock range |
-40℃~+150℃ |
||
Temperature fluctuation |
≤±0.5℃ |
|||
Temperature deviation |
≤±3℃ |
|||
Transform time |
≤15S |
|||
Temperature revert time |
≤5min |
|||
Temperature revert conditions |
sample load:10kg |
Sample exposed 30min in 150℃, 5min in ambient temp.,30min in -55℃, |
|
|
Capacity |
100 L |
|||
Workroom dimensions (mm) |
500D*500W*400H |
|||
Exterior dimensions (cm) |
115D*195W*210H |
|||
Power supply |
AC380V,50Hz,3 phase and 4 wire + grounding wire |
|||
B. STRUCTURE
C. CONTROLLER
D. REFRIGERATION SYSTEM
E. ACCESSORIES
F. SAFETY DEVICES
G. FACILITIES
Payment Terms: | TT/LC |
---|---|