Model: | AR-325 |
---|---|
Brand: | ARI |
Origin: | Made In China |
Category: | Electronics & Electricity / Other Electrical & Electronic |
Label: | Anaesthesia machine , Medical Equipment , Anesthesia machine |
Price: |
-
|
Min. Order: | 1 pc |
Last Online:29 Aug, 2016 |
AR-325 Anesthesia Machine
Main body |
High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Scope of application |
adult, child |
Gas source |
O2: 0.27--0.55MPa N2O: 0.27--0.55MPa |
Flow meter |
O2 : 0.1--1.0L/min 1.1--10L/min N20: 0.1--1.0L/min 1.1--10L/min Air : 0.1--10L/min |
O2, N2O linkage and N2O Stopper |
When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off. |
Flow rate of rapid oxygen supply |
25--75L/min |
Low oxygen pressure alarm |
There will be sound alarm when the oxygen pressure < 0.2MPa |
Vaporizer |
It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0--5vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
Respiratory circuit |
working mode: all-close, semi-close, semi-open APL: 0.5--7kPa |
Respiratory Bellows |
bellows for adults, tidal volume range:0--1500 ml bellows for children, tidal volume range:0--300 ml anyone can be chosen for application as required by the customer |
Ventilator |
|
Display mode |
High-definition 8.4” TFT color LCD screen display |
Ventilation mode |
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH(IMV, VCV, PCV) |
Ventilation parameters |
Tidal volume 50--1500 ml (20--1500 ml) Rate 1--99 bpm SIMV rate 2--20 bpm I:E 2:1--1:8(4:1--1:8) Inspiratory trigger pressure -1.0--2.0kPa PEEP 0--2.0kPa Pressure range 1.0--6.0kPa SIGH 1.5 times the inspiratory time (70-150 adjustable) Inspiratory platform 0--50% |
Parameters for ventilation monitoring |
tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate-time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance, flow-volume loop, pressure-volume loop) |
Monitoring of oxygen concentration |
21%--100% |
Security Alarm System |
|
Oxygen Concentration alarm |
upper limit setting range 21%--100% low limit setting range 10%--80% |
Airway pressure alarm |
upper limit setting range 0.3--6.0kPa low limit setting range 0.2--5.0kPa |
Per-minute ventilation volume alarm |
upper limit setting range 3.0--30L/min low limit setting range 1.0--10L/min |
Sustained high-pressure alarm |
it will give alarm when stress have consistently been higher than 2.5kPa |
Suffocation alarm |
it will give sound and light alarm if there is no tidal volume input for 15 seconds |
Power alarm |
|
Power |
AC 220V±10% 50Hz±1Hz(UPS, Storage battery) |
(Mechanical Arms) |
|
(End-expiratory CO2 monitoring) |
|
*Items within brackets are optional. |