Model: | AE-S70-1U |
---|---|
Brand: | A & E Lab |
Origin: | Made In China |
Category: | Electronics & Electricity / Electronic Instrument / Laboratory Instrument |
Label: | spectrophotometers , lab instrument , photometer |
Price: |
-
|
Min. Order: | 1 pc |
A & E Lab.
Main Products: Atomic Absorption Spectrometer (AAS), Fluorescence Spectrophotometer, Bio-Spectrophotometer, UV Vis Spectrophotometer, Visible Spectrophotometer, Flame Photometer, Whiteness Meter, Turbidity Meter, Colorimeter, Gloss Meter, Tintometer, Automatic Polarimeter, Auto Melting Point Apparatus, Automatic Refractometer, Abbe Refractometer, PH/DDS/TDS Meter, etc.
Introduction:
Key Features:
* Single beam spectrophotometer with highly stable optics.
* Extremely versatile with full functionality UV range: 190 to 1100nm, VIS range:320-1100nm.
* Spectral bandwidth 1.8 & 2nm.
* 6 inches & compact LCD display. Integrated user interface.
* USB port for data storage.
* ISO/CE Certificate passed.
* A & E Lab PC software available.
* Extensive range of accessories available.
Application:
AE-S70 Series spectrophotometers have the most attractive cost performance. It is widely used in colleges and enterprises for general
quantitative analysis and experiments.
* The instrument use a lead screw structure so that the instrument wavelength accuracy and wavelength resolution can be greatly improved.
* 8mm thick optical base: the instrument use a rigid die-cost aluminum base as its optical mount to ensure instrument
stability and reliability.
* It can set up various standard curves according to customer/s solutions and find the concentration of unknown solutions.
Ordering Information:
Order Code |
Description |
AE-S70-1U |
Large screen, single beam, 4 cell holder, 190-1100nm, Bandwith: 1.8nm. |
AE-S70-1UPC |
Large screen, single beam, 4 cell holder, 190-1100nm, Bandwith: 1.8nm., with PC software. |
AE-S70-2U |
Compact screen, single beam, 4 cell holder, 190-1100nm, Bandwith: 2nm. |
AE-S70-2UPC |
Compact screen, single beam, 4 cell holder, 190-1100nm, Bandwith: 2nm, with PC software. |
AE-S70 Series Visible Spectrophotometer |
|
AE-S70-2V |
LCD display, 320-1100nm, bandwith: 2nm, wavelength accuracy: ±0.5nm. |
AE-S70-2VPC |
LCD display, 320-1100nm, bandwith: 2nm, wavelength accuracy: ±0.5nm, with PC software. |
Technical Specifications:
Wavelength |
Photometric |
|||
Wavelength Range |
UV Range: 190-1100nm Visible Range: 320-1100nm |
Photometric Accuracy |
±0.3%T |
|
Wavelength Accuracy |
±0.5nm |
Photometric Repeatability |
0.2%T |
|
Wavelength Repeatability |
±0.3nm/±0.2nm |
Photometric Display Range |
0-200%T, -0.3-3.0A, 0-9999C |
|
Band Width |
1.8nm/2nm |
Stability |
0.002A/h @ 500nm |
|
Others |
Baseline Flatness |
±0.002A/±0.0015A |
||
Data Output Port |
USB |
Noise |
±0.001A |
|
Printer Port |
Parallel Port |
Stray Light |
≤0.05%T @220nm, 360nm |
|
Optical System |
Single Beam (1200 lines/mm grating) |
|||
Display |
LCD (320X240/128X64 Dots ) |
|||
Detector |
Silicon Photodiode |
|||
Lamps |
Deuterium & Tungsten Halogen Lamp |
|||
Power Supply |
AC 220V/50Hz, 110V/60Hz. |
|||
Packing Size & Weight |
680X590X420mm & 23kg |
A & E Laboratory Co.Ltd | |
---|---|
Country/Region: | Guang Dong - China |
Business Nature: | Manufacturer |
Phone: | 15800016110 |
Contact: | Stella Choi (Sales) |