Technical Parameters of Infusion Pump JSB-1200 |
Specifications |
Type of product |
JSB-1200 |
Detector |
outside caliper rotary detector (patent technology) |
Control way |
drop counting and automatically adjust the infusion rate |
Compatible IV set |
compatible with any brand IV set produced by any company(include lucifugal IV set) |
Alarms |
Air-in-line, infusion complete when volume limit has been set, empty container, occlusion, door opening, over time waiting, no drop been detected, error operation, low battery, device failure. |
Delivery rate range |
1.0ml/h~1200ml/h(1.0ml/h~99.9ml/h, 0.1ml/h step; >99.9ml/h, 1ml/h step) |
Delivery time range |
1~1999min (1 min step) |
Infusion volume range |
0~9999ml (1 ml step) |
Purge rate |
600 ml/h |
KVO rate |
when delivery rate ≤10ml/h, KVO is 1 ml/h ; Otherwise KOV is 3ml/h |
Accuracy |
≤25 drop/min: ±2%, >25 drop/min: ±3% (Rate mode and Time mode), ±5%(Volumetric mode with correct calibration) |
Bolus |
Bolus rate:10-200 ml/h Bolus volume: maximum of 100 ml |
Power Supply |
AC power supply: AC 100 to 240 V +/_10%, 50/60Hz Internal lithium battery: 11.1V, 2000mAh. 8 hours of continous operation on a new fully charged battery with infusion rate of 20 drops/min |
Power comsumption |
≤25 VA |
Safety level |
class Ⅰ, type CF |
Dimension & weight |
130mm(L)X 200mm(W)X230mm(H) |
Net weight |
approximately 2.5kg |
Operation conditions |
ambient temperature: 5 ~40 degree C Relative humidity: 20 to 90% |
Storage conditions |
ambient temperature: -10 ~ 55 degree C Relative humidity: 10 to 90% |
Quality Standard |
comply with EN-ISO13485(2003), IEC60601-1, IEC6 0601-2-24, ISO9001: 2000, ISO9001: 2008 |
Accessory |
AC power line, operation manual |