 I. Introduction

ETCR148 Super-large Caliber Clamp Leakage Current Sensor is well designed and manufactured for measurement of AC leakage current, leakage current, higher harmonic current, phase, frequency, power and electric energy, by adopting the latest CT technology. Its large caliber 148mm×108mm can clamp electric cable of Φ108mm, or 160mm×4mm flat cable and steel earth wires. It is portable, clamp designed, no need to disconnect the circuit, non-contact measurement, which means safer and faster. It can be connected with phase detection analyzer, industrial control equipment, data recorder, oscilloscope, electric power quality analyzer, high precision multi-meter, widely applied in those fields as electricity, communications, meteorology, railroad, oilfield, construction, measurement, scientific & research teaching institutes, industrial and mining establishments.

The clamp is made of special alloy, adopt the latest CT and magnetic shielding techniques, can almost shield the influence from external magnetic field, to ensure the high precision, high stability and high reliability of perennial uninterrupted measurement.

 

. Technical Specifications

 Function

Measurement of AC current, higher harmonic current, phase, electric energy, power and power factor

Test Mode

Clamp CT

Clamp Size

148mm×108mm

Measurement Range

AC 0.0mA-3000A

Resolution

AC 0.1mA

Accuracy

±2.0%FS (50Hz/60Hz; 23±2, below 70%RH, keep the measured wire be in the center of the clamp)

Coils Turn

3000: 1

Phase Error

≤3°(50Hz/60Hz; 23±2)

Reference Load

RL: 0-300mA≤3kΩ; 0-30A≤30Ω; 0-4000A≤0.3Ω

Output Mode

Current induction output

Output Interface

2 bare wires or standard multi-meter pen plug

Lead length

2m

Electric Field Interference

About 5mA when the external electric field 100A, 10mm nearby

Wire Position

Keep the measured wire be in the center of the clamp

Current Frequency

45Hz-60Hz (Frequency of the measured current)

 

10Hz-100kHz

Circuit Voltage

Below 600V AC

Size

Length 350mm × Width 180mm × Height 55mm

Weight

1.5kg

Working Temperature and Humidity

-20-50, below 80%rh

Storage Temperature and Humidity

-10-60, below 70%rh

Insulation strength

AC3700V/rms (between core and shell)

Safety Specifications

IEC1010-1, IEC1010-2-032, 2 class of pollution, CAT (600V)

Accessories

Current Sensor: 1 pc

 

ETCR148 Super-large Caliber Clamp Leakage Current Sensor 1