Total Aflatoxin - Precise Rapid Test

Total Aflatoxin - Precise Rapid Test
Model:-
Brand:Quicking
Origin:Made In China
Category:Industrial Supplies / Animal Husbandry Equipment
Label:total Aflatoxin , test , Aflatoxin kit
Price: -
Min. Order:1000 pc
Inquire Now

Product Description

1. INTENDED USE

Quicking total Aflatoxin - Precise Rapid Test is a competitive immunoassay for the semi-quantitative detection of the presence of Aflatoxin residue in feed or grain sample. 

Cut-off:   5 ppb base

Assay Time:  5 - 10 min

Sample: Feed or grain extraction

2. PRINCIPLE OF THE ASSAY 

Quicking total Aflatoxin - Precise Rapid Test is based on competitive lateral flow immunochromatographic assay. The Aflatoxin conjugate in the test zone will capture the immuno-gold (colloid gold-AF antibody conjugate), when there is very little dissociative AF in the extraction. A visible red test band indicates a negative result when the control band (C zone) shows that the card is valid. The test band (T zone) will be not visible if AF is present in concentration of the cut-off value and above which explains a positive result. 

3. KIT COMPONENT

- 10×foil pouches, each containing one TAF cassette with a pipette and a desiccant.

- 1×assay diluent (50 mL)  

- 2×centrifugal tubes (15 mL)

- Products Manual

4. TEST PROCEDURE

- Grind a representative sample to the particle size of fine instant coffee (50% passes through a 20 mesh screen). Weigh out 2.0 g ground portion of the sample into the centrifugal tube (15 mL).

- Add 2 mL of water and 8 mL of ethyl acetate into the tube. Shake for 5 min. (If emulsification appears, do a centrifugation at 4000 rpm for 1 min.)

- Collect accurately 2 mL of supernatant (ethyl acetate layer) into a small container (beaker or tube). Dry the liquid by blowing wind. 

- Redissolve the residues in the small beaker with an amount of the provided assay diluent. (REF: Table ) 

Cut-off (ppb)

5

10

20

50

100

Assay Diluents (mL)

0.4

0.8

1.6

4.0

8.0

- Take out the cassette from the foil pouch and place it horizontally.

- Suck the liquid and drip 3 drops into the assay sample hole “S”.

- Interpret the result in 5-10 minutes. Result after 10 minutes is considered as invalid.

5. INTERPRETATION OF RESULTS

Positive: Only one clear band in C zone indicates a positive result. Positive shows that the concentration of Aflatoxin is at or above 5 ppb in the samples.

Negative: The presence of both bands in C zone and T zone. 

Invalid: No band appears in C zone.

6. STORAGE

The kit can be stored at room temperature (2-30°C). The test kit is stable through the expiration date (18 months) marked on the foil pouch. DO NOT FREEZE. Do not store the test kit in direct sunlight.

Total Aflatoxin - Precise Rapid Test 1Total Aflatoxin - Precise Rapid Test 2

Member Information

Quicking Biotech Co.,Ltd
Country/Region:Shang Hai - China
Business Nature:Manufacturer
Phone:50947188
Contact:Yohey Sun (International sales)
Last Online:18 Feb, 2019