  • Scope of Application

      This product is mainly used in researching the VOC emissions change rule from wood-based panel and other light industry products. According to the results, the detected products can be graded, and environment quality can be evaluated.

     This product meets the standards of GB18580-2001,GB18584-2001,

GB18587-2001, ASTM D6007-02, EN 717-1, EN13419-1, ASTM D5116-97 ,

ASTM D6330-98,ASTM D6670-01,ASTM E1333-96.

  • Standards

GB18580—2001  

Indoor decorating and refurbishing materials-Limit of formaldehyde emission of wood-based panels and finishing products

GB18584-2001

Indoor decorating and refurbishing materials-Limit of harmful substances of wood based furniture

GB18587—2001

Indoor decorating and refurbishing materials-Limitations of harmful substances emitted from carpets,carpet cushions and adhesives

ASTM D6007—02

Standard Test Method for Determining Formaldehyde Concentrations in Air from Wood Products Using a Small-Scale Chamber1

EN 717—1    

 Wood-based panels-Determination of Formaldehyde Release

 Part1: Formaldehyde Emission by the Chamber Method

EN13419-1    

 Construction-Products Determination of VOCs Release

Part1:Emission Method of Environmental Testing Chamber

 

Reference Standards:

ASTM D5116-97

Standard Guide for Small-Scale Environmental Chamber Determinations of Organic Emissions From Indoor Materials/Products


ASTM D6330-98 

Standard Practice for Determination of Volatile Organic Compounds (Excluding Formaldehyde) Emissions from Wood-Based Panels Using Small Environmental Chambers Under Defined Test Conditions

             
ASTM D6670-01

ASTM D6670-01 Standard Practice for Full-Scale Chamber Determination of Volatile Organic Emissions from Indoor Materials/Products


ASTM E1333—96

 Standard Test Method for Determining Formaldehyde Concentrations in Air and Emission Rates from Wood Products Using a Large Chamber

 

3.Technical Parameters

1 Temperature Range:  (1030)℃    Precision±0.5℃

2 Humidity(3080)%RH.  Precision±3%R.H.

3 Air Exchange Rate:0.2 m32 m3/h(Rate: 0.22pcs/h   PrecisionVentilation quantity:±1%

4 Air Speed On Surface Air speed on level maintain by0.10.3m/sPrecision≤0.05m/s

5 Interior Dimension:(1±0.02m3

6 Interior Pressure10±5Pa

7 Tightness At static state1KPaLeakage rate≤1×10-3 m3/min

8 Average power:1.5KW Rated power:3KW

9. Protection

1High voltage protection ,Over-current protection of refrigeration system

2Over-temperature protection of heater

3Over-current protection of electrical machine

4Lack-water protection

5Over-heating,over-current,over-pressure protection of air pump

6Out of phase protection, Phase order protection

7Over-current protection,Short-circuit protection of power

10.Environment condition

1Temperature5℃35℃

2Humidity≤85%

3Pressure86kPa106kPa

4No pollution sources

11.External DimensionW1120mm×H1350mm×L1950mm

 

 