Fixed Oxygen gas detector under high temperature

Fixed Oxygen gas detector under high temperature
Model:OC-F08
Brand:Oceanus
Origin:Made In China
Category:Electronics & Electricity / Electronic Instrument / Analysis Instrument
Label:High temperature , gas detector
Price: US $500 / pc
Min. Order:1 pc
Inquire Now
点击这里给我发消息

Product Description

Description:

OC-F08 series online gas detector used the original imported sensor and the ultra low power 32 bit microprocessor, so with the advantages of high precision, good repeatability, stable signal, respond quickly, etc. Possible connected with many controller system, such as PLC, DCS, computer and other, with the function of remote monitor and alarm, data logger and analysis.

Common applications is gas safety inspection and alarm for industrial scene for Oxygen gas inspection under the high temperature working place.
Features:
4-20mA standard analog signal and RS485 digital signal output, 200~1000HZsignal output, relay output, data restore, data logger(optional);  
LED screen displaying the gas concentration, gas type, unit, instrument status etc.
Independent air chamber, the sensor is not required to replace the sensor calibration, automatic identification of key parameters of the sensor
Protection function to prevent high gas concentration attack.

Display the gas concentration in ppm, %LEL, %vol and mg/m3 is selectable.

Infrared remote controller operation;

With the 16bit ADC acquisition chip, designed with four-layer circuit board, so with high anti-interference ability.

With the Ex-proof certificate, the Ex-proof grade is Exd II CT6.

Technical Parameters:

Applicable Gas

Oxygen (O2)

Detecting Range

0-5%Vol、0-25%Vol、30%Vol、100%Vol or customized

Resolution

0.01%VOL, 0.1%VOL

Working principle

Zirconia

Operating Environment

Temperature: 0℃~700℃ (sensor working temperature)

Humidity: ≤95%RH (non-condensing); if >90%RH, can add the filter

Pressure: 86kPa~106kPa

Output Signals

(Standard)Three-wire system 4~20mA standard signals output, available for PLC, DCS, etc;

(Optional) Four-wire system RS485 digital signal output;

(Optional) Frequency signal 200~1000Hz output;

Two-level relay output:220VAC 0.5A, 5VDC 0.5A;

Installation Methods

Wall mounted, Pipe type(Thread:M40*1.5 )

Explosion-proof Index

Ex d IIC T6

Protection Grade

IP66

Carol

Henan Oceanus Import&Export Co.,Ltd.

Payment Terms:TT/Western Union
Fixed Oxygen gas detector under high temperature 1Fixed Oxygen gas detector under high temperature 2Fixed Oxygen gas detector under high temperature 3Fixed Oxygen gas detector under high temperature 4

Member Information

Henan Oceanus Export&Import Co.,Ltd.
Country/Region:He Nan - China
Business Nature:Manufacturer
Phone:13673683950
Contact:Carol Liu (Sales)
Last Online:01 Mar, 2019