Product Features |
|
|
Medical & ITE safety approvals |
|
2MOPP input to output isolation |
|
Suitable for medical equipment up to class BF(∧) |
|
Low Leakage current≤ 10μA (8.1-14V), ≤ 50μA (6.1-8V&14.1-24V) |
DOE efficiency level VI |
|
|
CoC V5 Tier 2(2016) |
|
|
Standby power ≤0.075W |
|
7V to 52V outputs, up to 24W |
|
Interchangeable AC plugs |
|
Protections against: short circuit/over current/ over voltage |
Technical Data |
|
|
Input Voltage:90~264VAC |
|
Input Current:0.5A |
|
|
Standby Power :0.075W |
|
|
Output voltage range:±5% |
|
Environmental Requirements |
|
Operationg Temperature: 0~40℃ |
|
Storage Temperature:- 20~60℃ |
|
Operating Humidity:10~90% |
|
Storage Humidity::5~90% |
|
Operating Altitude:Up to 5,000 M operating altitude |
|
Safety Approvals |
|
|
Safety aAgency/Mark |
Medical |
I.T.E |
CB |
IEC60601-1 |
IEC60950-1 |
UL |
ANSI/AAMI ES60601-1 |
UL60950-1 UL62368 |
CAN/CSA C22.2 NO.60601-1 |
CAN/CSA C22.2 NO.60950-1 |
TUV |
EN60601-1 |
EN60950-1 |
RCM |
- |
AS/NZS 60950.1 |
Mechanical Details |
|
|
88(L)*30(W)*57(H)mm |
|
|
Applications: |
|
|
Medical Device and ITE Equipment |