Model: | CS‐320 |
---|---|
Brand: | Yanrui |
Origin: | Made In China |
Category: | Electronics & Electricity / Electronic Instrument / Laboratory Instrument |
Label: | carbon sulphur analz , CS analyzer , CS320 |
Price: |
-
|
Min. Order: | 1 pc |
CS-320 Carbon / Sulfur Analyzer
General information:
Chongqing YanRui Instrument Co.,Ltd is one of the leading manufacturers of analytical instrumentation in China. YanRui is committed to providing customers with high-quality scientific instruments, professional services, and efficient C/S and O/N/H analysis and test solutions.YanRui keeps optimizing and improving product lines based on the orientation of customer and market demands. The product application covers metallurgy, foundry and casting, mechanical processing, electric power, mining, petrochemical, security and other fields industries.
CS-320 High-frequency infrared carbon/ sulfur analyzer adopts high-frequency combustion and infrared detection technology to analyze the content of carbon and sulfur in inorganic materials. Detection limit: carbon up to 100.00%, sulfur up to 100.00%. The CS-320 can be used in all metal and non-metal production and processing industries and meets laboratory-level requirements from enterprise and academic research.
Application Ranges:
Ferrous metals, non-ferrous metals, alloys, carbides, rare earth metals, ores, cements, ceramics, carbon compounds, coal, coke, catalysts, lime, gypsum, soil, glass and other solid raw materials.
Features
Specifications
Measured elements |
carbon, sulfur |
Field of application |
Ferrous metals, non-ferrous metals, rare earth metals, catalysts, alloys, ores, coal, coke, cement, ceramics, refractories, other metals and non-metallic solid materials. |
Function principle |
High-frequency furnace heating, infrared absorption detection |
Typical analysis time |
10s to 40s after combustion begins, depending on the sample analyzed |
Analytical range |
carbon:0.00001%~8% (Scalable to 100%) |
sulfur:0.00001%~3.5%(Scalable to 100%) |
|
Minimum readout |
0.000001% |
Analytical precision (repeatability) |
carbon:0.0001% or RSD≤0.5%(Meet one of them) |
sulfur:0.0001% or RSD≤1.0%(Meet one of them) |
|
Balance precision |
0.001g(optional 0.0001g) |
High frequency furnace |
Power:≤ 2.8KW, Oscillation frequency:18MHz |
Chemical required |
Magnesium perchlorate, sodium hydroxide |
Dimension |
W650mm×D790mm×H920mm |
Weight |
About 150 KG |
Function Principle:
CS-320 Polar adopts high-frequency furnace heating, infrared absorption detection systems. The samples are heated in oxygen condition, causing carbon and sulfur in the sample to react to CO2 and SO2. At the same time, trace CO is generated. The generated oxide is purified or catalyzed and passes into the infrared detection system. Using CO2 and SO2 at 4.26mm and 7.4mm, where there is a strong characteristic absorption band, the absorbed infrared light intensity is measured to calculate the percentage of carbon and sulfur elements in samples.