  MS-CJ-50C MS-CJ-100C MS-CJ-200C MS-CJ-300C
Interior capacity(L) 50 100 200 300
System 2-zone or 3 zone system by means of damper switching
Performance High temp. Chamber Pre-heat temp. Range +60℃~+200
heating time1 A+60℃→+200℃≤20min
Low temp. Chamber Pre-cool temp. Range -78℃~0
Cooling time1 A+20℃→-75℃≤70min A+20℃→-70℃≤80min
Test chamber Temp. uniformity ±2
Temp. Range A-65℃~+150 A-55℃~+150
Temp. Recovery time2 Within 5 minutes3 Within 5 minutes3 Within 5 minutes3 Within 5 minutes3
Speciment capacity 30kg
Speciment weight 2.5kg 7.5kg 10kg 7.5kg
Interior dimension(mm) W 410 650 650 970
H 460 460 460 460
D 370 370 670 670
Exterior dimension(mm)  W 1470 1710 1710 2030
H 1800 1800 1800 1800
D 1570 1570 1870 1870
Utility requirements Temp.+5℃~+35℃;Humidity:≤85%RHAtmospheric pressure86106kPa
Material Enclosure Galvanized steel plain Sheet color coated
Interior Stainless steel plate SUS304
Thermal insulant Hard urethane foam + glass wool
Construction High temp. Chamber Heater Nichrome Electronic Heater
Fan Centrifugal fan for High, ambient temperature.axial flow fan for pre-heating.
Low temp. Chamber Heater, condenser Nichrome Electronic HeatercondenserRegenerator
Fan Centrifugal fan
Drivers Air cylinder Drive the ventilation door
Air compressor Provide compressed air to drive the air cylinder (Optional)
Refrigerator unit Refrigeration mode Mechanical cascade refrigeration system(water cooled)
Refrigerating Compressor Semi-closed refrigerating compressor
Refrigerant R404a/R23
Condenser Stainless steel brazed plate heat exchanger
Controller Displayer 86.37(Height)×115.17(Longth)mmLCD
Running mode Program operation
Setting  Touch panel method
Program capacity 32 patterns, 3 steps max. per pattern, 999 times max. cycling.
Setting range High temperature upper limit:220℃;Low temperature limit-80
Test chamberShock upper limit+10℃,shock lower limit-10
Display resolution Temp0.1℃;Time0.1min
Input thermocouple
Control mode PID controlBTC Blanced Temperature control
Power supply AC380(1±10%)V (50±0.5)Hz three-phase four-wire+ protection ground wire
Power(KW) 38 36 36 36
Weight(kg) 1430 1480 1420 1530
Cyclic cooling water Water pressure0.250.4MPa Water temperature:≤30
Compressed air 0.50.7MPa
Standard accessories Test chamber status indicator 1pctime recorder 1pccable hole(Φ40)1pcCasters with adjusters.
Safety devices Leakage breakerInterior over heat protectorFan over heat protectorCompressor over heat protectorCompressor over pressure protectorSpeciment over heat protectorpower supply protector
1 Heating time and cooling time is measured under high temperature chamber  or low temperature chamber working only.
2 Recovery condition: Ambient temperature is +25 and cycling water temperature +25
  High Temperature chamber Low Temperature chamber Sensor position
3 Recovery time for high and low temperature chamber (+150℃:30minutes) (-65℃:30min) Upper side of speciment