Dynamic Test
Non equilibrium state test
Determination of dynamics constant and adsorption capacity
Isotherm PCT
Ideal balance state test
The temperature and pressure relationship determination
Acquirable thermodynamic properties
Cycle life test
Cycle PCT and life test
Adsorption and analysis test
Test influence on cycle adsorption, analysis and impurity
Hardware Features
Temperature control
The blue area of the instrument above keeps constant temperature around 30°,sample cylinder can be controlled from room temperature to -400°
Software features
Main Characteristic
Reference cylinder-equipped with a high pressure sample cylinder with random, 300ml volume and other volumes optional
Pressure range: vacuum-10MPa
Pressure control: PID Auto Control
Pressure detection: two pressure sensor, respectively is 50PSI and 1500PSI, accuracy to 0.05%FS,
Sample cylinder: multiple sample cylinders optional
ST100-304 stainless iron, temperature resistance 450°,pressure resistance 120MPa (standard)
ST110-cooper, temperature resistance 400°, pressure resistance 20MPa (optional)
ST120-Low temperature sample cylinder with low temperature producer (optional)
ST130-304 stainless iron, micro sample cylinder, mg grade sample adsorption (optional)
Safety
When hydrogen detector warns, gas circuit will be cut off automatically
As the switch main valve, pneumatic valve doesn’t use solenoid valve to ensure the safe
Pneumatic part is completely separated from electrical part
Vent is linked to the outer
Structure
Tube circuit—304 stainless iron
Pneumatic port--- NPT1/4' fitting
Size---length*width*height(45*40*50cm)
Net weight---40kg
Other accessories
Vacuum pump----secondary sliding-vane rotary vacuum pump
Air pump--------silent air pump
Connection pipe-----stainless iron pipe, O.D 3mm
Data collection card-----PCI high speed collection, COM port or Ethernet Interface
Power
AC 220V 50-60Hz 5A
Electric description and electric circuit in other countries and areas optional