single channel Syringe Pump LC1101 993.9ml/h
Product Description
Main Features for LC-1101 Syringe Pump:
- Antibolus feature
- KVO(Keep Vein Open)
- Near-end alarm to meet clinic needs
- With bolus rate(purge rate) function
- Adopt double MPU to ensure system security
- Occlusion alarm pressure is selectable
- Automatically recognize the syringe type
- Temporarily clear up alarm function and timing to recover
- Steamless-convert power supply system with internal battery embeded
- Pressing on/off button for 3 seconds is needed to turn off to avoid mistaken operation
- No changing the data during operation to avoid mistaken operation.
Detaled Spec.
Model No. |
LC-1101 |
Channel |
Single Channel |
Syringe Selection |
10 ml, 20ml, 50/60 ml |
Highest Infusion Rate |
99.9 ml/h |
Infusion Rate |
0.1~99.9ml/h(20ml/50ml/60ml) 0.1~99.9ml/h(10ml) |
Purge/Bolus Rate |
200 ml/h (20ml/50ml/60ml),99.9ml/h(10ml) |
Infusion Volume |
0-999.9ml, step by 0.1 |
Display Volume |
0-999.9ml |
Accuracy |
±3% |
Infusion Rtae Step |
0.1ml/h |
Alarms |
Syringe installation failure,Occlcusion,Near end,Pause overtime,Low-battery,Completion, Power interruption, Motor malfunction, Incorrect specification of the syringe,Electrical circuit jam. |
Occlusion Pressure:High |
20 ml,30ml,50/60ml syringe spec:90 Kpa±30Kpa; 10 ml syringe spec:100Kpa±30Kpa |
Occlusion Pressure:Low |
20 ml,30ml,50/60ml syringe spec:50Kpa±30Kpa; 10 ml syringe spec:70Kpa±30Kpa |
KVO Rate |
0.1ml/h±0.02ml/h |
Pause |
AC:100V~240V 50Hz/60Hz DC:12V |
Pause Overtiime Alarm |
1 min 50s~2min |