Model: | GDGY |
---|---|
Brand: | GOLD |
Origin: | Made In China |
Category: | Electronics & Electricity / Other Electrical & Electronic |
Label: | Automatic Insulating , Capacitance & Resist , Insulating Oil Tange |
Price: |
US $550
/ pc
|
Min. Order: | 1 pc |
Last Online:22 Apr, 2021 |
Features
Introduction
The Oil Tan Delta (OTD) is a laboratory instrument, which measures the Tan Delta, Resistivity and Relative Permittivity of insulating liquids. These tests give a wider picture of the overall quality of the insulating oil, and will indicate if there is any contamination.GOLD has developed features of the OTD to ensure that productivity levels within laboratories are maximized.
The microprocessor controlled inductive heater means that there is uniformity of temperature throughout the cell, in compliance with IEC 60247, and for cooling a fan effectively speeds up the cell cooling time. The instrument has a temperature measuring range of 10 to 110 ºC, with an accuracy of ± 1 ºC. It can be set to preheat before testing takes place. The automated oil drain facility negates the need to move the oil vessel.
User safety is paramount and with this in mind, GOLD have designed independent and dual redundant high voltage cut-off circuitry. If for any reason the user needs to abort the test, this can be done so by pressing any key, immediately terminating the test and removing the high voltage.
The design of the cell components maximise oil fill rate but minimise turbulence in oil flow and eliminate trapping of air.
With the calibration checker, you can carry out a quick measurement check to validate Tan Delta, Resistivity and Permittivity measurements. An on-site service with an Approved Service Centre for calibration provides the opportunity to adjust measurements where necessary and minimise downtime.
Specifications:
Item Name |
GDGY Automatic Transformer Oil Tan Delta Tester |
Condition of Use |
-5℃-40℃ RH<80% |
Power Source |
AC 220V±10% Without Frequency Limit |
AC High Voltage Output |
400V-2200V ±2% Every 100V 50VA |
DC High Voltage Output |
00V-2200V ±2% |
Temperature Control Induction Stove |
Maximal Power 500W |
Temperature Control Range |
<100℃ |
Temperature Measuring Error |
±0.5℃ |
Control Temperature Error |
0.1℃ |
Control Temperature Time |
Room Temperature 90℃ <20min |
Measuring Range |
tgδ Without Limit / Cx 15PF-300PF / Rx 10M-10T |
Resolution |
tgδ 0.001% / Cx 0.01pF/ Rx 0.01 |
Precision |
delta tgδ:±(Reading*0.5%+0.020%) delta Cx :±(Reading*0.5%+0.5PF) delta Rx :±Reading * 10% |
Relative Dielectric Loss Constant |
Εr Calculated Automatically |
Volume Resistivity |
ρ Calculated Automatically |
Dimension |
450(L)×310(W)×360(H) |
Net Weight |
21KG |