Model: | YG-301 |
---|---|
Brand: | sun |
Origin: | Made In China |
Category: | Chemicals / Coatings, Varnish & Painting |
Label: | - |
Price: |
-
|
Min. Order: | - |
Last Online:17 Nov, 2016 |
YG310 is a automatic closed flash point tester, can rapid test flash point of coating, paint (watebased paint), varnish, lacquer base, adhesive, solvent, Fatty acid methyl ester (FAME), chemical reagent, flavor, spices, solid chemical products, viscous chemical products, , air fuel oil, petroleum and related products.
Function Overview
Instrument can detect the flash point at a specific temperature between 0 ℃ and 100 ℃ (-30 ℃ 300 ℃ optional) or at average rate heating. All functions are achieved through the membrane keypad and digital display. Users can select the builin test procedures or set is own test parameters, you ca modify the default program to implement custom nonstandard test. Various paameters, such as the heating ode, time, temperature and the average rate heating can be set by the user themselves. Planning board action, the decline of the ignition, the detection of flash point is done automatically. A voice is prompted to user at each operation, at the end of the test there will be voice prompts, too. The gas ignition sources can be provided directly by the laboratory, you can also use the portable butane gas tank. Built-in cooling module is used to shorten the cooling time between the two tests.
Technical Parameters
1. Operate Conditions:
Operate Temperature: 15℃~35℃
Electric Supply: 220/240V,50/60Hz
2. Specifications:
Testing Method Meets Standards: ISO 3679, ISO 3680, ISO 1516 and ASTM D3278. ASTM D3828; and meets China Standards: GB/T520 -2008, GB/T21790-2008, GB5207 and GB7634.
Operate Method: Manually
Temperature Range: -0~100℃(-30℃~300℃ Optional)
Testing mode: Automatically testing (flash fire probe)
Testing Accuracy: 0.5℃
Sample volume: 2ml (4ml)
Atmospheric pressure: Auto fix
Testing mode: Rapid balanced mode and average rate heating mode
Heating Rate: Customize, quickest speed 5℃/min, longest duration 99min, highest temperature 100℃ (300℃)
Programming Test: The testing settings can be set through programming. Customer can modify and customize the settings by using built-in test program.
Cooling Method: Built-in cooling module (without external cooler 0℃~100℃)
Output Interface: Computer RS232, printer (optional)
Size: 400×220×170mm
Weight: 6kg
3. Back-up Spare:
Standard: Steam source regulation component
Silicone tube
2ml injector
Thermometer 0~100℃
O-Type Seal Ring (5)
Optional: Thick object injector
Powdery object injector
4ml injector
Flash point of quasi-standard oil (49.7℃)
Temperature Measurement Verifier (with certificate)
Ordering Information: YG-301(-0~100℃)
YG-301B(-30~100℃)
YG-301A(Room temperature~300℃)