|
Basic accuracy of CKT3544 series DC resistance tester is 0.02%, minimum resolution is 0.1μΩ, maximum test current is 1A, and the measurable range is 0.1μΩ~3.3MΩ. The test current is increased by the probe used to protect the port to achieve anti-interference measurement. Use the comparison judgment lamp (option) and the judgment sound of a large volume to accurately judge whether it is qualified or not at the production site. With EXT I/O interface suitable for NPN/PNP, it can be used in various automatic production lines. |
SPECIFICATIONS |
Model |
CKT3544 |
CKT3544-12 |
CKT3544-24 |
No. of Channel |
1 |
12 |
24 |
Parameters |
DC resistance |
Measurement Range |
0.1μΩ~3MΩ |
Min Resolution |
0.1μΩ |
Signal Source |
1A DC, Maximum DC 5.5 V |
Basic Accuracy |
0.1%±10 reading (3mΩ,30mΩ,3MΩ),0.05%±4 reading (300kΩ), 0.02%±2 reading (other ranges) |
Range |
3mΩ/30mΩ/300mΩ/3Ω/30Ω/300Ω/3kΩ/30kΩ/300kΩ/3MΩ |
Speed |
Fast (50Hz: 21ms, 60Hz: 18ms), slow (200ms) |
Temperature Compensation |
Range: -10℃~60℃, Accuracy:±1℃ |
Display Max |
32,000 |
Adjustment |
Short-circuit clear zero for all ranges |
Comparator |
10 bins sorting, output HIGH/IN/LOW sorting signal |
Trigger |
Internal trigger, External trigger, Manual trigger, BUS trigger |
Interface |
External IO interface, LAN interface, RS232/485 interface |
Others |
Temperature compensation function, comparator (ABS/REF%), lock button (OFF/menu lock/all lock), power frequency setting (AUTO/50Hz/60Hz), judgment tone setting, auto hold, average, panel save /read |
GENERAL FEATURES |
Power supply |
AC100~256V, Frequency: 50/60 Hz, rated power: 15 VA |
Size |
325mm (L)x 215mm (W)x96.5mm (D); Weight: 3.5kg |
ORDER ING INFORMATION |
Accessories |
KelvinTest leads; Temperature compensation probe;
RS232 communication cable; AC power cord |