Product Description

NCS-TT105 smart temperature transmitter

NCS-TT105 smart temperature transmitter,using the fieldbus technology, is a new generation of smart fieldbus temperature transmitter and it is an indispensable field device for process control.

NCS-TT105 transmitter integrates abundant function blocks and realizes not only general measurement function but also complicated control strategy.

NCS-TT105 uses digital technology, so it can connect with many types of thermocouple and thermo resistive sensors. It has wide range and simple interface between field and control room, which reduces the expense of installation, operation and maintenance.

NCS-TT105 supports HART, FF, and PA protocols. It can be widely used in the petroleum, chemicals, electricity, and metallurgical industries, etc.

Basic Parameters 

Input Signal

Resistance: PT100, CU50, CU100, 0~500Ω, 0~4000Ω

Thermocouple: B, E, J, N, K, R, S, T

Voltage signal: -100mV~100mV

Channels

2 Channels

RTD connection

2, 3 wire

Power Supply of Fieldbus

FF/PA:9~32 VDC/Current : ≤ 14mA

HART:12~42VDC

Insulation

Between housing and terminal board:

500 Vrms (707 VDC)

Display

6-digit numeric and 5-digit alphabet LCD Display

Operation Temperature

-40 ~ 85°C (Without display)

-30 ~ 70°C (With display)

Humidity Range

0% ~ 100% RH

Start Time

≤ 5s

Update Time

0.2s

Protection grade

IP 65

EMC

Designed to comply IEC 61000 (GB/T 18268-2000)

RTD Accuracy (25°C) 

Sensor Type

Working Range(°C)

Accuracy

Resistance

0~500Ω, 0~4000Ω

±0.05%

PT100

-200 ~ 850°C

±0.2°C

PT1000

-200 ~ 850°C

±0.1°C

CU50

-50 ~ 150°C

±0.3°C

CU100

-50 ~ 150°C

±0.2°C

RTD Other Specification  

Wiring

2, 3 wire

Update Rate

≥ 1HZ/CH

Common mode rejection

>80dB(50Hz)

Series mode rejection

>60dB(50Hz)

Temperature effect

<50ppm/°C

TC accuracy (25°C) 

Sensor Type

Working Range(°C)

Accuracy

Voltage

-100mV ~ +100mV

0.05%

B

500°C ~ 1810°C

±1.0°C

E

-200°C ~ 1000°C

±0.4°C

J

-190°C ~ 1200°C

±0.4°C

K

-200°C ~ 1372°C

±0.4°C

N

-190°C ~ 1300°C

±0.8°C

R

0°C ~ 1768°C

±1.0°C

S

0 °C~ 1768°C

±1.0°C

T

-200°C ~ 400°C

±0.4°C