electrosurgical generator

electrosurgical generator
Model:ECO-800
Brand:ECO
Origin:Made In China
Category:Electronics & Electricity / Other Electrical & Electronic
Label:electrosurgical unit
Price: US $1500 / pc
Min. Order:1 pc
Inquire Now

Product Description

Functions

This instrument is indicated for the cut and coagulation surgery of tissues in surgeries

 

Specifications

Voltage:AC 220V 50Hz

Maximum output power

Mode

Rated output power

ECO-800A

350W

ECO-800B I

300W

ECO-800B II

200W

ECO-800B III

100W

Working frequency :512 kHz/300kHz 

CF type equipment

Input power:600VA+10%

Unit size: 40cm * 42cm * 16cm

Weight: 18kgs

 

 

Characteristics

.Cut output power:0-350W, coagulation output power: 0-200W

.Six cut modes : High Power cut (ECO-8000A), Pure,Blend1,Blend2, Blend3, Endo cut

.Four coagulation Modes: Forced,Soft, Spray, Bipolar

.Monopolar and Bipolar auto-exchange

.Peak Power System(PPS) : Auto adjust power

.Spray coagulation is good for superficial coagulation.

.Wider Impedance Recognition(50-1800Ω), meeting all kinds of operation environment

.Neutral electrode Safety system(NESSY), continuously monitor neutral electrode

.Less smoke under power.

ENDO-CUT available to work with different coagulation mode

 

.Application

 

General surgery, orthopedics, gynecologic, urological,cardiac surgery, anus intestines, tumor surgery department and etc.

Endoscopic surgery under endoscope, celoscope and cystoscope;

Neurosurgery, microsurgery, E.N.T and Plastic surgery department under low power

 

.Model parameters

Mode

 Matching Impedance

Basic Parameters

Product Model

ECO-800A

ECO-800BI/D

ECO-800BII/D

ECO-800BIII

Cut

 

 

 

 

 

 

High Power

 

 

 

400Ω

Power

350W

300W

 

/

 /

Basic frequency

300KHZ

300KHZ

Pure

Power

300W

250 W

200W

100 W

Basic frequency

300KHZ

300 KHZ

300 KHZ

300 KHZ

Blend1

Power

150W

150 W

150 W

80W

Basic frequency

512KHZ

512 KHZ

512 KHZ

300 KHZ

Blend2

Power

120W

120 W

120 W

60W

Basic frequency

512KHZ

512 KHZ

512 KHZ

300 KHZ

Blend3

Power

80W

80 W

80 W

/

Basic frequency

512 KHZ

512 KHZ

512 KHZ

Endo

 

Power

120W

120

80 W

/

Basic frequency

512 KHZ

512

512 KHZ

Bipolar cut

250Ω

Power

350W

/

 

/

/

Basic frequency

100KHZ

Coagulation

 

 

 

 

 

 

Soft

 

400Ω

Power

150W

150 W

150 W

/

Basic frequency

512KHZ

512 KHZ

512 KHZ

Forced

Power

120W

120 W

120 W

100 W

Basic frequency

512KHZ

512 KHZ

512 KHZ

300 KHZ

Spray

Power

100W

100 W

100 W

/

Basic frequency

512KHZ

512 KHZ

512 KHZ

Bipolar coa.

  50Ω

Power

80W

80W

80 W

80 W

Basic frequency

300KHZ

300KHZ

300 KHZ

300 KHZ

 

Effect for cutting modes:

1- High cut: the model of 300W and above the energy , the model for underwater vaporization electric resection, can achieve the desired cutting effect.

 

2- Pure cut: similar to the effect of scalpel cutting , apply to fine and non-invasive cutting of any tissues.

 

3- Blend 1 cut : cutting along with mild clotting effect. For orthopedic surgery, flap separation, large blood vessels around the tissue separation.

 

4- Blend 2 cut : cutting along with moderate coagulation. Applicable to general, chest, laparoscopic cut, bleeding and other surgery.

 

5- Blend 3 cut: cutting along with stronger coagulation effect than the blend 2 cut . Applicable to the demolition of the breast, general surgery, orthopedics and so on.

 

6-  ENDO cut: in the endoscopic conditions, the operator couldnt directly control the operation process and set up a program called "ENDO CUT", refer to the actual clinical needs, flexible to set the output power of electric cut and coagulation, to achieve a more ideal surgical results.

 

7- Bipolar cut: unique in the saline , the cutting effect is caused by the Kinetic Energy Ion generated in Bipolar Vaporization, which can achieve the effect of rapid removal of tissue at low temperature, to avoid the occurrence of congestive edema and water poisoning.

 

8-Soft coagulation: Touched hemostasis, coagulation,are conducive to the wound healing after surgery. Suitable for rapid touched coagulation in luminal surgery and other fine tissue surgeries.

 

9-Forced coagulation: touched hemostasis, coagulation, Electron-coagulation effect is stronger, usually called point coagulation.

 

10-Spray coagulation: surgical electrodes and biological tissue can be without touching , apply to stop bleeding of a large area . The Electron-coagulation effect is strong, usually called surface coagulation (working with argon gas controller can produce argon ions for coagulation).

 

11-Bipolar coagulation: for neurosurgery or ordinary bipolar coagulation surgery. To achieve the tissues or vascular coagulation function in the middle two tweezers.

 

12-Bipolar Electro-coagulation: unique in the saline,the electro-coagulation effect is caused by the Kinetic Energy Ion generated in Bipolar Vaporization,

which can achieve the effect of rapid hemostasis and Coagulation at low temperature, to avoid the occurrence of congestive edema and water poisoning.

electrosurgical generator 1electrosurgical generator 2

Member Information

Nanjing kaihong healthcare co.,ltd
Country/Region:Jiang Su - China
Business Nature:Manufacturer
Phone:13371915
Contact:Rainy Day (sales)
Last Online:20 Apr, 2016